Bare Fiber Oxygen Microsensor for Customized Measurement Set-Ups
Implantable Oxygen Microsensor IMP-PSt7
This optical oxygen microsensor can be applied in various customized applications. The tiny probe has a tip size of 250 ?m with an outer diameter of 1000 ?m. The microsensor tip is not mounted into any additional housing. The bare plastic fiber tip can be mounted to your own housings, steel tubes and micro respirometer chambers etc. The small outer diameter even allows insertion into implantable Venflon-tubes. Moreover the microsensor can be deployed in sealed containers to measure the oxygen content directly. The IMP-PSt7 has a measurement range of 0 – 50 % oxygen, with a limit of detection of 0.03 % O2.
- High spatial resolution (250 μm)
- High temporal resolution (t90 < 3 sec.)
- Integration into customized housings and set-ups
- No consumption of oxygen
- Measurement in headspace & liquids
Technical
| Specifications | Gaseous & Dissolved O2 | Dissolved O2 |
|---|---|---|
| *at + 20 °C, after two-point calibration **TF = flat sensor tip with 250 ?m diameter | ||
| Measurement range | 0 – 50 % O2 0 – 500 hPa | 0 – 22.5 mg/L |
| Limit of detection | 0.03 % oxygen | 15 ppb |
| Resolution | ± 0.01 % O2 at 1 % O2 | ± 0.005 mg/L at 0.4 mg/L ± 0.025 mg/L at 9.06 mg/L |
| Accuracy at + 20 °C* | ± 0.05 % O2 or ± 3 % rel. | |
| Measurement temperature range | from 0 to + 50 °C | |
| Response time (t90) | TF**: < 3 sec. (gas) | TF**: < 10 sec. (liquid) |
| Properties | ||
| Compatibility | Aqueous solutions | |
| No cross-sensitivity | pH 1 – 14 | |
| Cross-sensitivity | Organic solvents, such as acetone, toluene, chloroform or methylene chloride | |
| Disinfection procedure | Disinfection with EtOH (70 % v/v) or isopropanol (100 % v/v) possible | |
| Cleaning procedure | Ethanol, soap solution, ddw | |
| Calibration | Two-point calibration with oxygen-free environment (nitrogen, sodium sulfite) and air-saturated environment | |
Related products
Resources
Publications
FAQs
Manuals
Oxygen Microsensors NTH / IMP / PM TF-250 (plastic fiber)
Oxygen Microsensors NTH / IMP / PM / FTCM (glass fiber)
Getting Started: Microx 4 & Microx 4 trace
PreSens Oxygen Calculator
Microx 4 & Microx 4 trace
Manual Micromanipulator MM
Automated Micromanipulator AM
User Guide: Automated Micromanipulator
Brochures
Software
PreSens Oxygen Calculator version 3.1.1 - Windows 7/8/10